documentation

Luminometer Standardization Kit (LSK)

The Luminometer Standardization Kit (LSK) is a linearity check for your luminometer.  Making sure that the luminometer produces a linear response is central to Luminultra’s product advantage of providing true total biomass concentrations rather than simply a generic machine output. 

Key terms

  • ATP: Adenosine triphosphate, energy molecule of living cells.
  • RLU: Relative Light Units, the measurement of luminescence.
  • RLU ATP1: Calibration signal from UltraCheck 1.
  • Background RLU: Baseline signal measured from Luminase only.

Run this test with the Relay™ app

Relay™ is your tool to easily calculate, organize, and share results. Access is included with your purchase.

Summarized instructions

If you are new to this test, please refer to the Full Method to follow the steps in detail.

  1. Log in to Relay™ . Set up a new site and sample location using the LSK test kit, then select RUN TESTS.
  2. Rehydrate Luminase with contents of the Buffer vial. Mix, and wait 5 minutes.
  3. Preparation: Add 100µL of Luminase to each of the 5 new test tubes.
  4. Measurements:
    • Add 2 drops of UltraCheck 0.01 to the first prepared test tube. Swirl and
      READ UltraCheck 0.1 RLU.
    • Repeat to measure the 4 remaining UltraCheck standards.
  5. Save: In Relay™, select SAVE TEST.

Component list

Kit componentStorage
Luminase enzyme & Buffer vials, 3mL20°C (68°F)
UltraCheck 0.01 dropper bottle, 2.5mL20°C (68°F)
UltraCheck 0.1 dropper bottle, 2.5mL20°C (68°F)
UltraCheck 1 dropper bottle, 2.5mL20°C (68°F)
UltraCheck 10 dropper bottle, 2.5mL20°C (68°F)
UltraCheck 100 dropper bottle, 2.5mL20°C (68°F)
200µL yellow pipet tips, 96 rack
12x55mm test tubes

1. Getting started

1.1 Log in to Relay™

If you are new to Relay™, follow the guides to first create a site and sample location.

  • Connect to your luminometer, then select RUN TESTS to begin a new test.
  • As you work through these test kit instructions, you can initiate RLU readings directly within Relay™ by selecting READ on any input. Once all inputs are entered, your results will automatically calculate.

1.2 Rehydrate the Luminase enzyme

  • Gently mix the Buffer and Luminase enzyme.
  • Wait 5 minutes for solution to dissolve.

After rehydration, keep Luminase refrigerated or frozen; it stays effective for 2–4 weeks at 4°C (39°F) or 3–6 months when frozen between uses. Before testing, let it thaw and reach room temperature (takes about 1 hour). Do not apply direct heat to warm it faster. For more details, visit my.luminultra.com.

1.3 Preparation

  • Use a new pipet tip to dispense 100µL of Luminase enzyme into each of the 5 new 12x55mm test tubes.

2. Measure ATP standards

UltraCheck 0.01

  • Add 2 drops (100µL) of UltraCheck 0.01 to the first prepared test tube.
  • Swirl the tube, place it in the luminometer, and take the UltraCheck 0.01 RLU reading (select READ) within 10 seconds.

UltraCheck 0.1

  • Add 2 drops (100µL) of UltraCheck 0.01 to the second prepared test tube.
  • Swirl the tube, place it in the luminometer, and take the UltraCheck 0.1 RLU reading (select READ) within 10 seconds.

UltraCheck 1

  • Add 2 drops (100µL) of UltraCheck 1 to the third prepared test tube.
  • Swirl the tube, place it in the luminometer, and take the UltraCheck 1 RLU reading (select READ) within 10 seconds.

UltraCheck 10

  • Add 2 drops (100µL) of UltraCheck 10 to the fourth prepared test tube.
  • Swirl the tube, place it in the luminometer, and take the UltraCheck 10 RLU reading (select READ) within 10 seconds.

UltraCheck 10

  • Add 2 drops (100µL) of UltraCheck 10 to the fifth prepared test tube.
  • Swirl the tube, place it in the luminometer, and take the UltraCheck 100 RLU reading (select READ) within 10 seconds.

3. Analysis

Calculations

Tip: Relay™ does the work For you

When all inputs are entered in Relay’s Run tests workflow, the results will calculate automatically. Select Save Test to catalog and share your results with your team.

In Relay™, confirm that the calculated R2 value is greater than 0.99 for good linearity.

If using Excel, generate a log-log graph with the obtained RLU values on the y-axis and the UltraCheck concentrations on the x-axis. Calculate R2 and confirm >0.99.

Data Interpretation Guidelines

Since Luminultra’s test kits use a single-point standardization (using only UltraCheck 1), a linear response is necessary. An R2 of 0.99 or greater is considered an acceptable result.

If the R2 falls below the 0.99 target, contact Technical Support for assistance.
 

Support

Troubleshooting

IssueRecommendation
Readings are lower than expected valuesEnsure reagents are not expired (expiry information can be found on reagent vials). Ensure reagents are being stored properly (storage information can be found on reagent vials)
Background readings are higher than expectedCheck if your Luminometer needs to be cleaned. Quarterly cleaning (LCK kit) and annual linearity verification (LSK kit) is recommended. Cleaning and maintenance kits are available for purchase from LuminUltra.

Additional Resources

myLuminultra:
https://my.luminultra.com

Contact support:
https://my.luminultra.com/hc/en-us/requests/new

Ordering Information

USA/CAN: +1 (506) 459 8777
UK/EU: +00 1 (506) 459 8777
Email: orders@luminultra.com