documentation

QuenchGone Organic Modified (QGOM)

The Quench-Gone™ Organic Matter (QGOM) Test Kit measures cellular ATP (cATP) to quantify total living biomass in hydrocarbon-based samples. By providing rapid, reliable insight into microbial growth, it helps you detect contamination early, prevent corrosion and fouling, and protect critical assets in energy and industrial systems.

Key terms

  • ATP: Adenosine triphosphate, energy molecule of living cells.
  • cATP (cellular ATP): ATP inside living cells, representing active biomass.
  • RLU: Relative Light Units, the measurement of luminescence.
  • RLU ATP1: Calibration signal from UltraCheck 1.
  • RLU cATP: Sample result for cellular ATP.
  • Background RLU: Baseline signal measured from Luminase only.
  • Volume: Amount of sample filtered (20–50mL).

Run this test with the Relay™ app

Relay™ is your tool to easily calculate, organize, and share results. Access is included with your purchase.

Summarized instructions

If you are new to this test, please refer to the Full Method to follow the steps in detail.

  1. Log in to Relay™ and select RUN TESTS at the site you are working with.
  2. Rehydrate Luminase with contents of the Buffer vial. Mix, and wait 5 minutes.
  3. Calibration: Mix 2 drops of UltraCheck 1 and 100µL of Luminase, and READ RLU ATP1.
  4. Filtration: Push up to 20mL of sample through the 20mL syringe + filter (record Volume).
  5. Wash & dry: Push 5mL of LumiClean through 20mL syringe + filter, then dry the filter using 50mL syringe.
  6. Extraction: Push 1mL of UltraLyse 7 through syringe + filter into an UltraLute tube. Cap and invert 3 times.
  7. Background: Add 100µL of Luminase to a clean test tube and read Background RLU.
  8. Assay: Add 100µL of UltraLute solution to a tube with 100µL of Luminase. Swirl and
    READ RLU cATP.
  9. Save: In Relay™, select SAVE TEST.

Sample volume recommendations

Sample TypeRecommended volume (mL)
Polymers, admixtures, personal or home care products1 (diluted)
Metalworking fluids, crude oil1–5
Oilfield, oily brines10–20

Component list

Kit componentStorage
Luminase enzyme & Buffer vials, 3mL20°C (68°F)
UltraCheck 1 dropper bottle, 2.5mL20°C (68°F)
UltraLyse 7 Bottle, 125mL20°C (68°F)
UltraLute (dilution) tube, 9mL20°C (68°F)
LumiClean bottle, 250mL20°C (68°F)
QuenchGone Organic syringe filters
20mL syringe, PP/PP
50mL syringe, PP/neoprene
100–1,250µL blue pipet tips, 96 rack
1–200µL yellow pipet tips, 96 rack
12x55mm test tubes
Disposable bulb pipets, 5mL

1. Getting started

1.1 Log in to Relay™

If you are new to Relay™, follow the guides to first create a site and sample location.

  • Connect to your luminometer, then select RUN TESTS to begin a new test.
  • As you work through these test kit instructions, you can initiate RLU readings directly within Relay™ by selecting READ on any input. Once all inputs are entered, your results will automatically calculate.

1.2 Rehydrate the Luminase enzyme

  • Gently mix the Buffer and Luminase enzyme.
  • Wait 5 minutes for solution to dissolve.

After rehydration, keep Luminase refrigerated or frozen; it stays effective for 2–4 weeks at 4°C (39°F) or 3–6 months when frozen between uses. Before testing, let it thaw and reach room temperature (takes about 1 hour). Do not apply direct heat to warm it faster. For more details, visit my.luminultra.com.

1.3 Calibration using UltraCheck 1 (RLU ATP1)

  • Hold the UltraCheck 1 bottle vertical and add 2 drops (100µL) of it to a 12x55mm test tube.
  • Pipet 100µL of Luminase into the tube.
  • Swirl the tube, place it in the luminometer, and take the RLU ATP1 reading (select READ) within 10 seconds.

If RLU ATP1 ≤ 5,000 rehydrate a new bottle of Luminase.

Once UltraCheck 1 is opened, it must be used within 3 months. After 3 months, discard and use a new bottle.

2. Cellular ATP analysis (RLU cATP)

2.1 Filtration

  • Mix sample well.
  • Remove the plunger from a 20mL syringe.
  • Attach a new syringe filter.
  • Pour the recommended volume of sample into the syringe, and record the amount used as Volume.
  • Place the plunger back into the syringe and slowly push sample through the filter into waste receptacle.
  • Detach the filter.
  • Remove the plunger.

2.2 Washing and drying

  • Re-attach the filter to the 20mL syringe barrel.
  • Using a disposable bulb pipet, add 5mL of LumiClean to the syringe barrel.
  • Push the LumiClean slowly through the filter into a waste container.
  • Remove the plunger from a 50mL syringe.
  • Detach the filter from the 20mL syringe.
  • Attach it to the 50mL syringe.
  • Hold the 50mL syringe over the waste receptacle and push the plunger through the barrel to dry the filter.

2.3 Extraction

This step breaks open microbial cells to release ATP.

  • Remove the plunger from the 20mL syringe.
  • Re-attach the filter to the 20mL syringe barrel.
  • Add 1 mL of UltraLyse 7 to the barrel using the 1mL pipettor.
  • Pass it slowly through the filter and collect in a new 9mL UltraLute (dilution) tube.
  • Cap and invert three times to mix.

2.4 Background measurement

  • Add 100µL of the Luminase into a new, clean 12x55mm test tube using the 100µL pipettor and a yellow pipet tip.
  • Put this test tube into your luminometer and take the Background RLU reading (select READ).
  • This tube can then be used for your first sample assay on the next step (Step 2.5).

If the Background RLU is above 20 RLU, this indicates that some external factors (light, contamination) may be influencing your results. LuminUltra offers cleaning kits for your luminometer. Please contact us to discuss.

2.5 Assay

  • Combine 100µL of the UltraLute (dilution) solution (containing your filtered, extracted, and washed sample) with 100µL of Luminase.
    • For the first sample, add the UltraLute (dilution) solution directly to the test tube you used for the Background measurement, using a new pipet tip.
    • For each subsequent sample, add 100µL of UltraLute (dilution) solution and 100µL of Luminase to a new 12x55mm test tube, using new pipet tips for each reagent and each sample.
  • Swirl the tube, place it in the luminometer, and take the RLU cATP reading (select READ) within 10 seconds.
  • In Relay™, select SAVE TEST.

3. Analysis

Calculations

Tip: Relay™ does the work For you

When all inputs are entered in Relay’s Run tests workflow, the results will calculate automatically. Select Save Test to catalog and share your results with your team.

\mathrm{cATP}\;(\mathrm{pg\ ATP}/\mathrm{mL})
=
\left[
\frac{\mathrm{RLU}_{\mathrm{cATP}}-\mathrm{RLU}_{\mathrm{background}}}
     {\mathrm{RLU}_{\mathrm{ATP1}}}
\right]
\times
\frac{10{,}000\ (\mathrm{pg\ ATP)}}{V_{\mathrm{sample}}\ \mathrm{(mL)}}

Data Interpretation Guidelines

Sample TypeGood Control
(pg cATP/mL)
Preventative
(pg cATP/mL)
Corrective Action
(pg cATP/mL)
Polymers, Admixtures, Personal Care, Home Care<100100–1,000>1,000
Crude Oil, Oily Brines, Chemical Products, Oilfield Waters<100100–1,000>1,000
Metalworking Fluids, Fire- Retardant Lubricants<1,0001,000–10,000>10,000

Our guidance thresholds (e.g., 10 pg/mL) are drawn from broad industry experience with microbial contamination in similar systems. We encourage you to build your own baseline by comparing ATP results with on-the-ground observations. This creates thresholds tailored to your operation—so your team can act with confidence in every decision.

Support

Troubleshooting

IssueRecommendation
Readings are lower than expected valuesEnsure reagents are not expired (expiry information can be found on reagent vials). Ensure reagents are being stored properly (storage information can be found on reagent vials)
Background readings are higher than expectedCheck if your Luminometer needs to be cleaned. Quarterly cleaning (LCK kit) and annual linearity verification (LSK kit) is recommended. Cleaning and maintenance kits are available for purchase from LuminUltra.

Additional Resources

myLuminultra:
https://my.luminultra.com

Contact support:
https://my.luminultra.com/hc/en-us/requests/new

Ordering Information

USA/CAN: +1 (506) 459 8777
UK/EU: +00 1 (506) 459 8777
Email: orders@luminultra.com